KXE-TM-BA_IA-e-1411

 Ensure that there are no objects on the weighing plate.

Wait for stability, then press

.

After approx. 5 seconds the zero point adjustment is completed.

Cancel the adjustment with

, the appliance returns into the menu.

 Having adjusted the zero point, „SPAn“ will be displayed, followed by the

currently set weight value for the adjustment weight.

 Either use the displayed adjustment weight or change it with

,

and

(numerical input see chapter 2.1.1), the active digit is flashing.

 Place the indicated adjustment weight and wait for stability.

Acknowledge with

.

After approx. 5 seconds the adjustment is completed.

• Cancel the adjustment with

, the appliance returns into the menu.

• Back to weighing mode, actuate adjustment switch.