SEA Series Steam Sterilizers

Operation manual

1 3

7. Printer User's Guide

According to the requirement of customer, Runyes steam
sterilizer can adopt prescriptive printer. Connect the printer
and the steam sterilizer with prescriptive cord. Switch on the
power, two lights on the printer Panel will go on, means
printer has been installed.

Caution: Runyes steam sterilizer can only select micro
printer offered by Runeys.

Print information explaining:

Date:Sterilize date

CN:name of cycle

C.S.T:The starting time of cycle

Vp1:vacuum pressure1

Pp1:The phase of pressure1

H.S.T:the starting time of sterilization.

H.E.T:the end time of sterilization.

D.S.T:the starting time of drying.

C.E.T:the end time of cycle.

Operator:Operator You can signature name according to

practicality instance

Caution:Chinses characters provided by Runyes.the

mini-autoclave of Runyes is only to equip

with the mini-printer of Chinese characters

provided by Runyes

Fig7-1

Sterilization

Records

Date : 07 . 04. 20
Cycle: 0000 1
C . N . : 1 3 4 2 . 1 0

C. S.T :23 : 15 :16
V p 1 : 2 3 : 2 1 : 5 5

0 . 0 5 0 4 6 . 6

P p 1 : 2 3 : 2 3 : 5 7

- . 8 0 0 5 0 . 1

H . S . T . : 2 3 : 5 0 : 2 4

M a x : 2 . 2 4 1 3 4 . 6

M i n : 2 . 1 4 1 3 4 . 5

Ave:2.19

134.5

H . E . T . : 0 0 : 0 8 : 2 7

D .S .T. : 00: 10 :44

0 . 3 0 1 1 8 . 5

C.E .T. : 00: 10 :45

Op e r at o r:

Resu lt: Passed

T ha nks