SEA Series Steam Sterilizers

Operation manual

2

2.Specifications

Chamber inner dimensions: (Diameter x Depth)-----------------------------------200Mm x 358mm
Power Supply:--------------------------------------------------------------------- According to the nameplate
Sterilizing Temperature/Pressure --121 /(1.0-1.3) Bar (Under standard atmospheric pressure)

--------------------------------------- 134 /(2.0-2.4) Bar(Under standard Atmospheric pressure)

Note: Temperature/Pressure here means the chamber temperature and pressure during the phase
of sterilization. The value varies under different atmospheric pressure.

Power Fuse -----------------------------------------------------------------------------------------------T12A
Water reservoir capacity-------------------------------------------------------------------------- 3.5 liters

Environment temperature -------------------------------------------------------------------- +5

+40

Relative Humidity Range -------------------------------------------------------------------------- 30% 75%

Figures

Package Key

Start/Stop key

Select Key

Dry Key

Test Cycle Key

Used Water Outlet

Cycle Step Key

Caution & Warning

Type B Device

Grounding

Tip:

Standard atmospheric pressure: It is the atmospheric pressure at sea level. It refers to 1
Standard atmospheric pressure. The value is 0.1Mpa=1000Hpa=1.0 Bar. Atmospheric pressure
decreases by higher altitude. For instance: at 3000 meters or lower, An increase of altitude of
1000 meters will result in a decrease of ATM PR .by 0.1Bar approximately.