SEA Series Steam Sterilizers

Operation manual

2 9

Important information

1.The autoclave must be installed on a leveled counter top.

2.When operated at a higher elevation of more than 500 meters,The machine must be
reprogrammed. Please get in touch with your distributor for instructions.

3.Distilled water must be used to prolong the life of the machine.

4.The radiation openings in the machine must not be blocked or covered.

5.Instruments should be placed on the tray allowing space between one another and thus
allowing steam to pass through.

6.Condensed water collector must be emptied from time to time.It should be emptied when
refilling the water reservoir.(Draining of the collector can be followed as per item 9.2)

7.Chamber door knob must be turned to its locking position.

8.Never try to open the chamber door before pressure displays 0.0 .

9.To avoid burning injuries, power off the machine and let it cool down before replacing the seal.

10.No dragging when the machine is moving.

11.Power connection must be grounded properly.

12.Must be provided enough power.(1800VA)

13.While the temperature of outside is lower, please pre-heating the machine without
instruments for 30minutes before sterilize.