SEA Series Steam Sterilizers

Operation manual

1 9

9.5Cleaning Steam generator

9.5.1 Aim: Clean the steam generator and steam pipe, to avoid
the furring block the pipe.

9.5.2 Description: 20 min cleaning + 10 min drying

9.5.3 Purpose: to clean the steam generator furring, take it out
of the machine through pre vacuum.

9.5.4 Operation: when the machine is inactivity, press SELECT
for 5 seconds, the B&D light and LEAK light will go on at mean
time, it means you chose the cleaning cycle. Close the door and
press START, it will start to clean.

9.5.5 Cleaning remind: after cleaning, every 100 cycles, B&D
light and LEAK light will go on at the mean time and glitter.
Remind you to clean the steam generator.

Friendly remind: Cleaning function will not affect the autoclave
cycle, only remind message.

Fig 9-10