Type 3 Commands – Balance Output Functions:

Command Response

Purpose

SEND[CR]

See the RS-232 Balance Message Formats section.

Returns the contents of the display register.

[register number]_RCL[CR] REG

XXX:[register

contents][current units or

'MULT.'] [CR] [LF]

Sends the contents of the specified register
in the current units if it is additive or with
the designator 'MULT.' if it is multiplicative.
An unflagged number is sent with no
designator. An undefined register returns a ?
with a [CR][LF].

RCL_TARE[CR]

REG 091:[TARE register contents] [current
units][CR][LR]

Return the TARE register's contents.

RCL_[(CONVERT)(CAL)] REG

XXX:[contents of specified

register][MULT.][CR][LF]

Returns indicated register contents, if in that
mode. If not in that mode then a ? with a
[CR][LF] is returned.

RCL_PCS[CR]

REG XXX:[contents of pieces register][last FP
units/pc][CR][LF]

Returns pieces register contents expressed as
last front panel units/pc, if in that mode. If
not in that mode then a ? with a [CR][LF] is
returned.

[register
number]_RCL_TARE[CR]

REG XXX:[register contents][current units][CR][LF] Adds the specified register contents to the

TARE register.

RCL_XAVG[CR]

[average weight] [current units] AVX[CR][LF]. [2
second delay]. [number of samples taken]
###[CR][LF]. [2 second delay]. [standard deviation]
[ current units] SDEV[CR][LF]. [2 second delay].
[relative standard deviation] COV[CR][LF].

Sends the results of a standard deviation run.
The TBAR mode, if in use, is temporarily
suspended while this process is going on.


Type 4 Commands – Clock/Calendar Functions:

Command Response

Purpose

[six digit integer]
_ENTER_100_STORE[CR]

Sets the calendar function, if available, to the
specified date. The date must be specified in
the MMDDYY format and no delimiters.

100_RCL[CR]

REG 100: MM.DD.YY DATE[CR][LF]

Returns the date.

[six digit integer]
_ENTER_101_STORE[CR]

Sets the clock function, if available, to the
specified time. The time must be specified
in the HHMMSS format using military 24
hour time format.

101_RCL[CR]

REG 101: [HH:MM:SS] TIME[CR][LF]

Returns the current time, using military 24
hour time format, once for each request.

[1 to 6 digit integer]
_ENTER_89_STORE[CR]

Sets the auto send register to the specified
send time interval. For the auto send feature
to be operational, you must set the auto send
selection to ON (see RS-232 Front Panel
Configuration section for details). The
format used is HHMMSS with no delimiters.
For instance, to set the auto send interval to
5 minutes and 25 seconds, you would send
this command string: 525 ENTER 89
STORE[CR]. Also you must have the auto
Send selection set to ON before the auto
Send will work (see RS-232 Front Panel
Configuration Section).

89_RCL[CR]

REG 089: [HH:MM:SS] INTV[CR][LF]

Returns the auto SEND time interval
currently set.

[(1)(0)]_ENTER_79_STORE[CR]

A 1 in register 79 disables the date/time
stamp feature of any SEND. A 0 in register
79 enables the date/time stamp.

Advertising