RS-232 Balance Message Formats:

The balance will respond to a SEND command by transmitting a message in one of three formats depending on how the SEND
command was issued. The following discussion defines four types of SEND commands and the three message formats.

1. The first case typically applies to a balance with RS-232 connected to a printer and an external print switch. If the SEND is

initiated by activation of the hardware SEND line on the RS-232 I/O module and the display shows the current weight, then the
response will be a format B. If the display does not show the current weight (i.e., date, time, or a register RCL), the hardware
SEND is ignored.

2. This case typically applies to a balance with the RS-232 interface connected to a printer. If the SEND is initiated by a time or

auto SEND and the display shows the current weight, then the response will be a format B. The date/time stamp is included in
format C if register 79 has been set to 1 prior to setting register 89 to the desired time interval. To stop auto sending set register
89 to 0 or send an ASOFF command. If the date/time stamp is enabled and the display doesn't show the current weight, the
date/time stamp is still sent. Nothing is sent if the date/time stamp is disabled and the display doesn't show the current weight.

3. This case typically applies to a balance with the RS-232 interface connected to a computer. There are several RS-232 I/O

commands containing the keyword RCL. These commands, issued by the computer, all cause a single format C or a sequence of
format C send to occur.

4. This case also typically applies to a balance with the RS-232 interface connected to a computer. A SEND command issued over

the RS-232 I/O channel, by the computer, causes an unconditional format A send.

Format A - Balance in the normal weighing mode:

1 2 3 4 5 6 7 8 9 10 11 CR
X X X X X X X - - - Y......

LF


X = a digit (0-9) or a decimal point or a space or a minus sign.

Y = an upper case alphabetic character (A-Z). The first character of the units or mode annunciator in the display, followed by the rest
of the characters of the annunciator.

If the display shows a minus sign, it is the first character in the balance message. In this case the least significant digit of the displayed
number is in position 8 followed by only two spaces instead of three. The 1st character of the units annunciator is still in the 11th
position.

Here are some examples. The notation used here is that CR means the carriage return or enter character; LF means the line feed
character; and _ indicates a single space character. Note that all space characters are shown explicitly.

Display shows 5.15 g
Message is

_ _ _5.15_ _ _G CR LF


Display shows 211.05 dwt
Message is

_211.05_ _ _DWT CR LF


Display shows -211.05 dwt
Message is

-_211.05_ _DWT CR LR


Display shows .0035 A
Message is

_ _.0035_ _ _A_SPEC. CR LF

Format A - Balance in a special weighing mode:


Display shows 1250 PCS
Message is

_ _1250_ _ _ _ _PCS CR LF


Note that, in these special weighing modes, the least significant digit of the numeric field is in the 6th (or 7th position if the number is
negative) and that the leftmost character of the annunciator is in the 12th position.

Display shows -100.00 CAL
Message is

-100.00_ _ _ _CAL CR LF


Display shows sigma Tbar 0.00 g
Message is

_ _ _ 0.00_ _ _G_SIGMA_ _TBAR_ CR LF