Troubleshooting guide – Scientech Zeta Series User Manual

Page 19

Advertising
background image

19

Troubleshooting Guide:


Problem

Possible Cause

Possible Solution

The display is blank

• Balance not turned on

• Power cable not plugged in

• No power from AC outlet

• Press ON/OFF switch

• Plug in cable

• Turn on circuit breaker

• Change AC outlets

OL appears in display

• Maximum capacity exceeded

• Reduce container weight

• Weigh sample in smaller increments

• Switch to high range (dual range units)

UL appears in display

• Pan is not in place

• Balance is out of range

• Insure pan is positioned properly

• Press ZERO switch

The display is unstable

• Drafts (air currents are present)

• Vibrations are present

• Install windscreen, relocate balance

• Isolate balance from vibration, relocate

balance

Weight readings are incorrect

• Balance is out of calibration

• Balance is not level

• Display was not zeroed before weight

was placed on pan

• Object being weighed is touching

windscreen or balance case top

• Unit of weight selection is incorrect

• Recalibrate balance

• Level balance

• Press ZERO switch before weighing

• Reposition object

• Select proper weighing mode

Display stays in turn on sequence

• Pan not in place

• Drafts or vibrations present

• New windscreen has been installed

• Power supply connected to wall outlet

before balance connector

• Place pan on balance

• Isolate or relocate balance

• Raise pan height

• Unplug power supply from wall then

re-plug into wall

Partially lit display, “weird”
characters displayed or “garbage”
on the display

• “Live” cord plugged into balance

• Unplug power supply from the

electrical outlet, wait 5 seconds, then
plug back in

Err appears in the display

• Procedural error

• Review procedure in manual and try

again

Err 1 appears in display

• Calibration weight is incorrect

• Use calibration weight that appears in

balance display

Err 2 appears in display

• Balance not warmed up enough for

calibration to be performed

• Plug in and turn on balance for 1 hour

Err 3 through 14 appears in
display

• Service is required

• Call Scientech's Service Department at

(800) 525-0522 or (303)444-1361 or
email us at [email protected].