ERROR MESSAGES
Displayed error Meaning
Out of error mode
Err01
Load cell signal is unstable
Automatic
Err02
Overload on full scale
Automatic after
removing weight
Err03
Gross negative. This action is not allowed
Automatic
Err04
Out of zero range
Press any key
Err05
Sampling accuracy too low
Press any key
Err06
Input signal too high
Automatic after
correcting input
Err07
Input signal too low
Automatic after
correcting input
Err08
Calibration out of range (negative)
Automatic
Err09
Calibration out of range (signal too low)
Automatic
Err10
Calibration count 2nd(3rd) point lower than
count 1st(2nd) point
Automatic
Err14
Set-point value 2 < set-point value 1. This is
not allowed
Automatic
Err98
Calibration point must be higher than
previous one
Automatic
Err99
Action only allowed in start-up units
Automatic