Triner Scale RAVAS-310 User Manual

Page 24

Advertising
background image

24

MULTI-POINT CALIBRATION

Push the T key for about 10 seconds.

After 3 seconds the display will show the last entered pre-set tare value.

After 7 seconds the display will go into the calibration mode.

The display will show the first calibration point with the pointer “e1” flashing.

Using the

and

keys you can see the three earlier programmed values on the display. The

pointer will move through e1-

3. “e1” is the first calibration point, “e2” the second and “e3” the third.

Use the

and

keys to return to the first point.

The indicator shows the value of the first

calibration point, with the “e1” pointer flashing.

Load the weighing system with a known weight.

Press the

key to enter this weight onto the indicator.

The first segment will start flashing.

Use the

,

and

keys to change all segments until the proper weight has been entered.

Press the

key to return to calibration mode.

The “e1” pointer will start flashing.

Press the

key for 3 to confirm the entered weight.

This calibration number counts down from Adj 08 to Adj 00, the first calibration point has now
been set.

Move to the second calibration point.

The display will show the pointer “e2” flashing.

Repeat the procedure for a second known weight. Be aware that the value of this weight has to be
higher than that of the first weight. If not, the display will show ERR98 and return to the entry
mode for the calibration point.

Repeat the procedure for the third known weight. Leave calibration mode by pressing the

or

key until AP XX appears.

This number indicates the calibration sensitivity percentage, eg AP 07.

Press the

key.

The display now shows the value of the gravitation constant. Use the

,

and

keys to

correct this for your position.

Press the

key to return to the standard weighing mode.