Triner Scale TS-700 Series V4.0 User Manual

Page 10

Advertising
background image

Calibration & Connectivity | TS-700 Series

8

calibration parameters, proceed as follows:

CALIBRATION PROCEDURE

1) Power on the indicator.

2) Press the PRINT and HOLD keys. “C” will

display. Release the keys.

3) “C01” will display, with the 1 flashing. The

indicator is now is Parameter Settings mode.

Navigating in Settings Mode

When the indicator is in settings mode, refer to

the graphic printed below the key to determine

the keypad key’s function.

The

and

arrow keys are used to

increase/decrease the value of the flashing digit,

or to navigate up and down a list of preset

options.

The

and

arrow keys are used to move the

active digit over to the left or to the right.

The

key and

key can be used to move

forward or backward through the parameters.

Also, the

key is used to enter/accept

changes after they have been made to the

parameter settings.

The

key can be used to exit settings mode.

With the indicator in settings mode and the

calibration weight nearby, proceed with the

calibration process while referring to the

following chart:

NOTE: After completing and verifying the

calibration, set the calibration switch to the “Off”

position and securely fasten the rear plate.