Calibration & Connectivity | TS-700 Series

7

Calibration Switch, TS-700 MS: Remove the

screws and the back plate of the enclosure. A

white-capped calibration switch is located at the

bottom left corner of the circuit board. Press the

switch fully in and release. This will place the

switch in the up position, allowing access to the

calibration parameters. Loosely reattach the

rear plate.

Calibration Switch, TS-700 SS: Remove the

screws and the back plate of the enclosure. A

calibration switch is located at the bottom left

corner of the circuit board. Move the switch to

the “On” position, allowing access to the

calibration parameters. Loosely reattach the

rear plate.

Calibration Switch, TS-700 WB: Disconnect all

cables and cords. To remove the back plate of

the enclosure, place the indicator upside down

(not face down) on a work surface. Locate the

two slotted holes near the connection sockets.

Insert a flat blade screwdriver into each slot

(one at a time), pushing the blade inward while

angled towards the back of the indicator. This

will release the retaining prong and enable

removal of the back plate. Carefully lift the rear

plate.

A white-capped calibration switch is located at

the bottom left corner of the circuit board. Press

the switch fully in and release. This will place

the switch in the up position, allowing access to

the calibration parameters. Loosely reattach the

rear plate.