Calibration & Connectivity | TS-700 Series

6

Calibrating a TS-700 Series Indicator

NOTICE: The TS-700 SS/MS/WB are

NTEP Class III Legal for Trade approved

and certified. Unlocking the calibration

parameters is not allowed by persons

other than qualified scale technicians.

ALL MODELS: Place the Indicator in

Calibration Mode and Determine if the

Calibration Parameters are Accessible

1) Power on the indicator.

2) Press the PRINT and HOLD keys. “C” will

display. Release the keys.

3) If the calibration parameters are accessible,

“C01” will display, with the 1 flashing. Proceed

to page 8, “CALIBRATION PROCEDURE”.

NOTE: If C08 Displays instead of C01, the

calibration switch must be adjusted. Disconnect

the AC adapter and power off the indicator.

DEFAULT SETTINGS CAUTION
Placing the indicator in calibration mode enables
access to the Default Settings Parameter, CO7.
NEVER activate Parameter C07 unless you are a
qualified scale technician,
and:
A) You understand that all custom settings will be
deleted permanently, and
B) You understand how to restore all custom
parameters to the required settings based on the
scale’s specifications and usage.