Error codes – Triner Scale TS-700 Series V4.0 User Manual

Page 17

Advertising
background image

Calibration & Connectivity | TS-700 Series

15

Error Codes

ERROR CAUSE SOLUTION

uuuuuu

1. Overloaded condition.
2. Wrong connection at

load cell.

3. Load cell has quality

problem.

1. Reduce the weight.
2. Check load cell

connection

3. Inspect load cell.

Check the input and
output.

nnnnnn

1. Calibration is incorrect.
2. Bad connections.
3. Load cell has quality

problem

.

1. Check scale is resisted

or not, foot is kept
level or not.

2. Check load cell

connection.

3. Check input and

output resistance of
load cell

.

ERR1

Overweight calibration
error.
Incorrect weight input or
incorrect weight on
weighing platform.

1. Input weight correctly

during calibration.

2. Place correct weight

on weighing platform.

ERR2

Underweight calibration
error.
Incorrect weight input or
incorrect weight on
weighing platform.

The calibration weights
Minimum is 10% of
Max. cap.
Recommend 60%-
80% of Max. Cap.

ERR3

During calibration, the
input signal is negative.

Check all connections.
Check load cell.
If connections and
load cell are okay,
PCB needs replacing.

ERR4

During calibration, the
signal is unstable.

After the platform is
stable, start
calibration.

ERR5

Circuitry error

Replace the PCB.