Allow sufficient cooling of the pump before

starting maintenance.

+

Vent the pump and isolate it from the

vacuum system before you start mainte-

nance. Drain condensate, if applicable.

Ensure that maintenance is done only by suitably

trained and supervised technicians.

The valves and diaphragms are wear parts. If the rated

ultimate vacuum is no longer achieved or in case of in-

creased noise level, the pump interior, the diaphragms

and the valves must be cleaned and the diaphragms and

valves must be checked for cracks or other damage.

All bearings are encapsulated and are filled with long-life

lubricant. Under normal operating conditions, the drive

system is maintenance free.

In demanding circumstances, it may be efficient to check

and clean the pump heads on a regular basis. In normal

use, the lifetime of the diaphragms and valves is typically

15,000 operating hours.

- Prevent internal condensation, transfer of liquids or

dust. The diaphragms and valves will be damaged if

liquid is pumped in significant amount.

- Carry out maintenance frequently if the pump is ex-

posed to corrosive media or in case of deposits.

- Regular maintenance will improve the lifetime of the

pump and also protect both users and the environment.

NOTICE

atm

°C / °F

Set of seals (diaphragms, valves and O-rings).........................................2x 696867

for

ME 16C NT VARIO / PC 3016 NT VARIO

Set of seals (diaphragms, valves and O-rings) ............................................696839

for

MD 12C NT VARIO, MV 10C NT VARIO / PC 3012 NT VARIO / PC 3010 NT VARIO

Diaphragm key (w/f 66) ...............................................................................636554

+

Service only one side of the pump at a time to avoid the mixing of

parts.

+

In case, support the pump appropriately.