Electrical connection, Notice – VACUUBRAND MV 10C NT VARIO User Manual

Page 43

Advertising
background image

page 43 of 145

+

Maximum permissible pressure at the pres-

sure transducer: 21.8 psi (1.5 bar) absolute.

+

Attention: At pressures above approxi-

mately 795 Torr (1060 mbar) the pressure

reading becomes incorrect due to satura-

tion of the pressure transducer. The display

will flash. Release pressure immediately!

Risk of bursting!

maximum

21.8 psi

(1.5 bar)

absolute

Condensate and deposits will affect the mea-

surement results.

Position the CVC 3000 and its vacuum line

in such a way that condensate cannot flow

towards the pressure transducer. Clean the

pressure transducer, if necessary. See section

“Cleaning the pressure transducer”, pg. 81.

NOTICE

Connect the CVC 3000 controller to the

VACUU•BUS cable of the NT VARIO pump.

Attention: Do not apply off-axis forces when

assembling or removing plug connections! Ob-

serve correct orientation of the plug. To con-

nect further components use VACUU•BUS Y-

adapters and extension cables.

Electrical connection

O

➨ I

Emission condenser Peltronic:

Plug in the power cord.

• Check the power source and the Peltronic’s

rating plate to be sure that the power source

and the equipment match in voltage, phase,

and frequency.

Connect the emission condenser Peltronic via the

VACUU•BUS cable to the CVC 3000 controller.

Switch the emission condenser Peltronic on.

Advertising