VACUUBRAND MV 10C NT VARIO User Manual

Page 53

Advertising
background image

page 53 of 145

The CVC 3000 controller can be adapted to the specific application by

choosing the appropriate function depending on the connected compo-

nents and the requirements of the application.
Automatic detection of the components

When switching on the controller, the configuration of the connected com-

ponents is checked automatically.

Connected components (pump NT VARIO, gauge head VSK 3000,

valves) are

detected automatically and controlled by the CVC 3000 un-

til the controller is switched off. Switch off/on the controller to renew the

configuration.

The last mode of operation and the preselected values (e.g., for pres-

sure, speed or time for switching off) are stored.

If the preselections are chosen appropriately, it is possible to start imme-

diately if similar operating conditions are desired.

The controller features

five functions and one configuration menu,

see section ”Menu guide”. Each of these functions involves different menu

options, which are presented automatically and reflect the connected

components.

The types of components connected (e.g., valves) de-

termine the active menu items.

Changing the function:

Switch controller on.

Press ”START/STOP” key to terminate control in case control is run-

ning (e.g., if ”Autostart” is activated).

Press ”MODE” key.

Select function with knob and press to confirm.

+

Depending on the selected function and system components, the con-

troller provides different operating control, as follows:

’’Pump down’’

• Manages the continuous speed control of the NT VARIO pumps de-

pending on preselected pressure and time settings.

• Coolant valve

Notes on selecting the function

Advertising