Technical data

General technical data valid for all pumps

Maximum permissible inlet pressure

(absolute)

psi

(bar)

16

(1.1)

Maximum permissible outlet pressure

(absolute)

psi

(bar)

16

(1.1)

Maximum pressure difference between

inlet and outlet

psi

(bar)

16

(1.1)

Permissible ambient temperature

storage / operation

°F

(°C)

14 to 140 / 50 to 104

(-10 to +60 / +10 to +40)

Permissible relative atmospheric mois-

ture during operation (no condensation)

%

30 to 85

Maximum permissible installation alti-

tude above mean sea level

ft

(m)

6500

(2000)

Rated motor power

hp (kW)

0.59 (0.44)

No-load speed 50/60 Hz

rpm

1500 / 1800

Maximum permissible range

of supply voltage (±10%)

Attention: Observe specifications of rating plate!

100-115 V~ 50/60 Hz,

120 V~ 60 Hz

230 V~ 50/60 Hz

Maximum start-up current

100-115 V~ 50/60 Hz, 120 V~ 60 Hz

230 V~ 50/60 Hz

A

A

15.5

6.1

Device fuse

2 slow blow fuses 8A

Motor protection

thermal cutout, manual reset

Degree of protection IEC 529

IP 40

Inlet

small flange KF DN 25

Outlet

silencer / thread G 1/2”

Dimensions L x W x H

approx.

in

(mm)

21.8 x 10.2 x 14.1

(554 x 260 x 359)

Weight approx.

lbs. (kg)

67.5 (30.6)

We reserve the right for technical modification without prior notice!