• Reduce the transmission of vibration. Pre-

vent mechanical load due to rigid pipelines.

Insert elastic hoses or flexible elements

as couplings between the pump and rigid

pipes.

Note: Flexible elements will compress or

flatten when evacuated if not designed for

use under vacuum.

• Hose connections at the pump inlet must al-

ways be gas tight.

Use a suitable adapter to connect a vacuum

hose (caoutchouc) 15 mm ID (see ”Accesso-

ries / Spare parts”, pg. 34), if necessary.

Use connecting hoses with large diameter and

keep them as short as possible to avoid flow

losses. Locate the pump as closely as possi-

ble to the application.

Secure hose connections at the pump ap-

propriately, e.g., with hose clamps, to protect

against accidental detachment.

Use a suitable valve (see “Accessories / Spare

parts”, pg. 34) to isolate the pump from the

vacuum application. This is to allow the pump

to warm up before pumping condensable va-

pors and to clean the pump after use before it

is switched off.