Connecting the outlet – VACUUBRAND MV 10 NT User Manual

Page 29

Advertising
background image

page 29 of 70

+

Particles and dust must not be aspirated. If necessary,

you must install appropriate filters. You must ensure

their suitability concerning gas flow, chemical resis-

tance and resistance to clogging prior to use.

• A power failure may cause accidental ventilation of the

pump. If this constitutes a potential source of danger,

take appropriate safety measures.

When assembling, ensure

vacuum-tightness. After as-

sembly, check the whole system for leaks.

NOTICE

Outlet via silencer.

Attention: Use the silencer only in case of low

gas throughput. Check the silencer regularly

for permeability! In case, install a hose nozzle

or a small flange connection instead (see “Ac-

“Ac-

Ac-

cessories / Spare parts“, pg. 34).

/ Spare parts“, pg. 34).).

Connecting an exhaust line at the pump outlet:

Unscrew the silencer and install a hose nozzle

DN 15 mm or a small flange connector KF DN

16 (thread G1/2”).

outlet

If dangerous or polluting fluids could be released at

the outlet, install an appropriate system to catch and

dispose of those fluids.

+

Connect a gas-tight exhaust line at the pump

outlet (see above) if necessary. Always vent

exhaust gases appropriately (e.g., into a

fume hood).

+

Never block the gas outlet. The exhaust

line must always be free of obstructions (no

back pressure) to ensure an unimpeded

discharge of gas. The cross-section of the

Connecting the outlet

NOTICE