Prevent internal condensation, transfer of liq-

uids or dust. The diaphragms and valves will

be damaged, if liquids are pumped in signifi-

cant amounts.

Check the pump regularly for external soiling

and deposits. Clean the pump if necessary

to avoid an increase of the pump’s operating

temperature.

In case of overload, the motor is shut down by

a

self-hold thermal cutout in the winding.

Note: Only manual reset is possible. Switch

off the pump and disconnect from the power

source. Identify and eliminate the cause of fail-

ure. Allow the pump to cool down sufficiently

before restart.

°C / °F

I ➨ O

!

3.

2.

1.

IN

Avoid overheating (e.g., due to hot process

gases). Maximum permissible temperature

range: see “Gas inlet temperatures”, pg. 24.

Do not start the pump if the

pressure difference between

inlet and outlet ports exceeds max. 16.0 psi (1.1 bar).

Attempts to start the pump at higher pressure difference

may cause stalling and damage of the motor.

NOTICE

Make sure the fan’s air supply is adequate. Do

not place the pump on soft surfaces (e.g., rub-

ber foam) during operation. This may reduce

the air circulation in the pump housing and af-

fect the fan’s cooling effect. Check fan regu-

larly for dust/dirt. Clean fan guard grill if neces-

sary to avoid a reduction of ventilation.

air

NOTICE

IN

condensate / dust

NOTICE

NOTICE

NOTICE