VACUUBRAND DCP 3000 User Manual

Page 50

Advertising
background image

page 50 of 64

Fault

Possible cause

Remedy

❑ Displayed pressure

value is fluctuating.

➨ Position of the pressure

transducer VSK 3000

has been changed (e.g.

from horizontal to verti-

cal)? Fluctuation of the

displayed pressure in

the range of 0.2 - 0.4

mbar?

✔ Readjust pressure trans-

ducer, if exact pressure

indication is required.

➨ Pressure transducer

VSP 3000 is still in the

warm-up phase?

✔ Allow for a warm-up time

of 20 minutes.

➨ Pressure transducer

VSP 3000 was ex-

posed to a significant

pressure change or a

change of position.

✔ Allow the pressure trans-

ducer to stabilize.

➨ Pressure measurement

with VSP 3000 at pres-

sures above 50 mbar?

✔ Fluctuations within the

range of the measure-

ment uncertainty. Use

pressure transducer VSK

3000.

❑ Pressure reading

incorrect.

➨ Connected pressure

transducer configured

incorrectly or several

pressure transduc-

ers with identical

VACUU•BUS address-

es connected?

✔ Reconfigure pressure

transducer.

❑ Second pressure

transducer con-

nected but not de-

tected by the DCP

3000.

➨ Second pressure trans-

ducer not or incorrectly

configured?

Both pressure trans-

ducers configured to

the same address?

✔ Configure second pres-

sure transducer.

❑ Display and warn-

ing triangle are

flashing.

➨ Data storage capacity

in data logger function

is reached?

✔ Read out data and start

again, if necessary.

A service manual with spare parts list and directions for repair is avail-

able on request.

+

The service manual is intended for trained service people only.

Advertising