Decontamination of the vacuum pump, Notice – VACUUBRAND BVC basic / control / professional /G: Instructions pour la décontamination User Manual

Page 13

Advertising
background image

page 13 of 27

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. 25/11/2013 // 999281

The vacuum pump of the BVC may contaminated due to operation. The protective

filter with a pore size of 0.2 µm does not avoid safely a contamination of the vacuum

pump in all cases.

Also in case of a defect of the filter or operation of the BVC without filter pathogenic

germs (e. g., bacteria, viruses, prions, fungi) may enter the pump.

A decontamination of the vacuum pump is possible in the following ways:

- Gassing of the interior.

- Disassambling the pump parts and insertion of the parts in a disinfectant solution

or by autoclaving of the contaminated parts.

Decontamination of the vacuum pump

only BVC control / G and BVC professional / G

The applicability of the decontamination method and the selection of the ap-

propriate disinfectant have to be ensured by the user.

NOTICE

NOTICE