VACUUBRAND BVC basic / control / professional /G: Instructions pour la décontamination User Manual

Page 22

Advertising
background image

page 22 of 27

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. 25/11/2013 // 999281

+

Position any washers that are present between diaphragm

support disc and connection rod.

Screw diaphragm clamping disc, diaphragm, diaphragm

support disc, and washers to connecting rod.

Bring the diaphragm into a position in which it is in con-

tact with the housing and centered with respect to the bore

so that it will become clamped uniformly between housing

and head cover.

Assemble head cover and valves.

+

Pay attention to the correct position of the guidance pin in

the head cover (circle in figure)!

+

Pay attention to

correct orientation of the valves (see

figure):

Inlet side of pump head (black valve): The valve tongue

points at the kidney-shaped orifice in the valve seat.

Outlet side of pump head (white valve): The valve is ori-

ented the opposite direction as the valve at the inlet side.

Round orifice under the valve tongue.

Lay down the BVC on the side of the pump, e. g., at the

edge of a workbench; support if necessary.

Put on housing cover and housing cover insert.

+

Move housing cover insert slightly to ensure that the head

covers are correctly positioned.

Pay attention to washers and position screws. Screw in

4 screws with washers at the housing cover diagonally,

loosely at first with a Torx driver T20, then tighten.

+

Do

not tighten until head cover is in contact with housing,

maximum torque: 2.2 ft

.

lb

f

(3 Nm).

Position hose to elbow fitting and screw on the union nut.

Position BVC upright.

Assemble the cover so that the groove in the in the face

side of the cover is at the inlet of the pump.