Notice – VACUUBRAND BVC basic / control / professional /G: Instructions pour la décontamination User Manual

Page 21

Advertising
background image

page 21 of 27

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. 25/11/2013 // 999281

Fill the hose to the vacuum sensor with disinfectant soluti-

on.

Empty the hose after the exposure time.

Decontaminate all disassembled parts and the used tools in an

appropriate disinfectant solution or by autoclaving.

Comply with exposure time.

Prior to reassembling let the parts dry completely.

If the device is send for repair disassembled, ensure that all parts are packed

safely.

Lift the diaphragm at the side. Place the diaphragm care-

Lift the diaphragm at the side. Place the diaphragm care-

fully together with diaphragm clamping disc and diaphragm

support disc in the diaphragm key.

+

Avoid damage of the diaphragm: Do not excessively bend

or crease the diaphragm.

Assemble the original number of washers between dia-

phragm support disc and connecting rod.

+

Too few washers: The pump will not attain vacuum spe-

cification. Too many washers: Diaphragm clamping disc

will hit head cover, causing noisy operation and possibly

causing the pump to seize up.

Position diaphragm between diaphragm clamping disc

with square head screw and diaphragm support disc.

+

Note: Position diaphragm with pale side towards dia-

phragm clamping disc (facing pump chamber).

+

Make sure that the square head screw of the diaphragm

clamping disc is correctly seated in the guide hole of the

diaphragm support disc.

Reassembling

Prior to decontamination clean surfaces with a damp cloth if necessary. Use water or

a mild detergent as cleaning agent.

NOTICE

NOTICE

NOTICE

NOTICE

Advertising