Notice – VACUUBRAND BVC basic / control / professional /G: Instructions pour la décontamination User Manual

Page 17

Advertising
background image

page 17 of 27

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. 25/11/2013 // 999281

Tools required (metric):

- Diaphragm key SW 46 (order no.: 637777)

- Screw driver Torx TX20

- Open-ended wrench SW 19

Decontamination of the pump parts in a disinfectant solution or by autoclaving

The vacuum pump has to be disassembled.

When disassembling avoid contact of contaminated parts with not contami-

nated parts.

The pictured pump parts have to be decontaminated.

The applicability of the decontamination method and the selection of the ap-

propriate disinfectant have to be ensured by the user.

Prior to decontamination clean surfaces with a damp cloth if necessary. Use water or

a mild detergent as cleaning agent.

NOTICE

NOTICE

NOTICE

Advertising