Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. 25/11/2013 // 999281

This manual describes possibilities to decontaminate the aspiration systems BVC

and the handle VHC

pro

.

Prior to decontamination clean with a damp cloth if necessary. Use water or a mild

detergent as cleaning agent.

Prior to sending the devices to the factory the devices have to be decontami-

anted completely.

The applicability of the decontamination method and the selection of the ap-

propriate disinfectants have to be ensured by the user.

Ensure that repair is done only by an appropriately trained and supervised tech-

nician. Obey local and national safety requirements. Ensure that the technician is

familiar with the safety procedures which relate to the products processed by the

pumping system. Ensure that the pump is decontaminated before repair and that

you take adequate precautions to protect people from the effects of hazardous sub-

stances if contamination has occurred.

+

Take adequate precautions to protect people from the effects of dangerous sub-

stances. Wear appropriate safety-clothing, do not inhale and avoid contact with
skin.

Prior to decontamination from outside and / or disassembling the device

isolate

equipment from mains. Allow sufficient cooling.

Ensure that the device cannot be operated accidentally. Never operate the de-

vice if covers or other parts are disassembled. Never operate a defective or

damaged device.

Wait 5 seconds before starting work to allow the capacitors to discharge.

Before starting check that the required parts (e. g. tools, disinfectnats) are available

and of the correct type.

Check the operating sequence mentally using exploded view drawings, spare parts

lists and circuit diagram on feasibility, safety requirements and consequences on

safety and function of the equipment.

Obey regulations when disposing of solvents and chemicals.

In order to comply with law (occupational, health and safety regulations, safety at

work law and regulations for environmental protection) vacuum pumps, components

and measuring instruments returned to the manufacturer can be repaired only when

certain procedures (see section

”Notes on return to the factory”) are followed.

Scrapping and waste disposal

Dispose of the pump and any components removed from it safely in accordance with

all local and national safety and environmental requirements.