Calibrate the sensor/transmitter:

Adjust the sensor/transmitter gain.

Apply appropriate calibration gas to the sensor.

When signal has stabilized, set the sensor/transmitter gain according to the instructions furnished with that device

Calibrate the MX52 C

ONTROL

:

3.

Third match the

MX52 C

ONTROL

display to the S/T display, set the value of the calibration gas on the

MX52

display panel by using the channel sensitivity potentiometer (Item A, Figure 11).

4.

Finally set the

MX52 C

ONTROL

current to the calibration output, adjust the current output by using the 20 mA

potentiometer to the appropriate output for the calibration gas being used (Item B, Figure 11).

Example of calibration current output: For a channel with a scale of 0 – 100 ppm:

If 100ppm gas is used to calibrate the S/T, the calibration current output should be set to 20mA

If 50ppm gas is used to calibrate the S/T, the calibration current output should be set to 12mA

Conclude Calibration of the MX52 C

ONTROL

:

Stop the injection of calibration gas, wait for the measurement to return to zero (on the

MX52

display panel).

Then, press the “CALIBRATION” key (Item C, Figure 10). The flashing yellow

LED

goes out and the “C” on

the display panel disappears. The channel now operates normally and calibration has been completed.

Figure 11: Potentiometer Layout on Front of Channel Board