Hanna Instruments HI 110 User Manual

Page 13

Advertising
background image

13

• If necessary, press the ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is

stable.

• When the reading is stable and close to the selected buffer, the

“READY” tag will be displayed and the “CFM” tag will blink.

• Press CFM to confirm calibration.

• If the fifth calibration point is reached, the instrument stores

the calibration value and returns to normal measurement

mode, or else it goes to the next calibration point.

TWO-POINT CALIBRATION
• Proceed as described in “THREE-/FIVE-POINT CALIBRATION” section.

• Press CAL after the second calibration point was confirmed. The instruments

will memorize the two-point calibration data and will return to measurement

mode.

ONE-POINT CALIBRATION
Two SETUP selectable options are available: “Pnt” and “OFFS”.

If the “Pnt” option is selected, the new calibration point overrides an existing

one. The adjacent slopes will be reevaluated.

If the “OFFS” option is selected, an electrode offset correction is performed

keeping unchanged the existing slopes.

• Proceed as described in “THREE-/FIVE-POINT CALIBRATION” section.

• Press CAL after the first calibration point was confirmed. The instruments will

memorize the one-point calibration data and will return to measurement

mode.

Note: • If the value measured by the instrument is not close to the

selected buffer, “WRONG” “ ” and “WRONG” “ ” tags will blink

alternately. In this case check if the correct buffer has been used,

or regenerate the electrode by following the Cleaning Procedure

(see page 30). If necessary, change the buffer or the electrode.

• When a custom buffer is displayed, the “BUFFER pH” tag blinks.

To change the custom buffer value in according with the buffer

temperature proceed as described in “WORKING WITH CUSTOM

BUFFERS” paragraph (see page 14).

• If the buffer temperature or the manual temperature exceeds the

temperature limits of the buffer, “WRONG” tag and temperature

reading will blink.

• If “WRONG”, “BUFFER pH” tags and “OLD” message are displayed

blinking on the secondary LCD line, an inconsistency between new and

previous (old) calibration is detected. Clear calibration parameters

and proceed with calibration from the current calibration point. The

instrument will keep all confirmed values during current calibration.