Hanna Instruments HI 110 User Manual

Page 26

Advertising
background image

26

CHR n

Change the instrument range according with the parameter

value (n):

• n=0 pH range/0.001 resolution

• n=1 pH range/0.01 resolution

• n=3 mV range

• n=4 Relative mV range

COMMANDS REQUIRING AN ANSWER
pH?

Causes the instrument to send the pH reading. If the range is set

to mV, “Err6” is received.

MV?

Causes the instrument to send the mV reading. If the range is set

to pH, “Err6” is received.

MR?

Causes the instrument to send the relative mV reading. If the

range is set to pH or mV, “Err6” is received.

TM?

Causes the instrument to send the temperature reading.

RAS

Causes the instrument to send a complete set of readings in

according with the current range:

• pH, mV and temperature reading on pH range.

• mV and temperature reading on mV range.

• Rel mV, absolute mV and temperature reading on Rel

mV range.

MDR

Requests the instrument model name and firmware code.

PAR

Requests the setup parameters setting (instrument ID, calibra-

tion alarm time-out, pH range, beep status).

GLP

Requests the calibration data record.

Note: • “Err8” is sent if instrument is not in measurement mode.

• “Err7” is sent if mV are asked during pH calibration mode.
• “NAK” (21) character is sent when the instrument receives an

unknown or a corrupted command.

Note: Invalid commands will be ignored.

Advertising
This manual is related to the following products: