Setup – Hanna Instruments HI 110 User Manual

Page 17

Advertising
background image

17

Setup mode allows viewing and modifying the following parameters:

• Calibration Alarm Time-Out

• Custom buffer 1

• Custom buffer 2

• One-point calibration behaviour

• Instrument ID (HI 112 & HI 113)

• Current Time

• Current Date

• Beep Status

• Baud Rate (serial communication, HI 112 & HI 113)

• Command prefix (serial communication, HI 112 & HI 113)

To enter SETUP mode press SET while the instrument is in measurement mode.

Select a parameter with the ARROW keys.

Press CAL if you want to change a parameter value. The selected parameter

will start blinking.

Press RANGE (SEL for HI 110) to toggle between displayed parameters.

Press the ARROW keys to increase or decrease the displayed value.

Press CFM to save the modified value or CAL to escape.

Press the ARROW keys to select the next/previous parameter.

INTV

CAL

CAL

or the number of days until the calibration alarm will be displayed.

or if expired (7 days ago).

• The instrument ID.

Note: • Press GLP at any moment and the instruments will return to

measurement mode.

• If calibration has not been performed, the instruments display

“no CAL” message blinking.

SETUP