RELATIVE mV CALIBRATION (HI 111

AND HI 113)

• Press CAL when the instrument is in RELATIVE mV measurement mode.

The “mV” and “ ” tags will blink. Absolute mV is displayed on the

primary LCD and “AbS” message is displayed on the secondary LCD.

• When the absolute reading is stable and in measurement range, the

instrument asks for confirmation.

• If the reading is out of range, “WRONG” tag will be displayed.

• Press CFM to confirm the absolute value. The instrument will display 0.0

mV on the primary LCD and “rEL” message on the secondary LCD. In this

moment the relative mV offset is equal to absolute mV reading.

• Use the ARROW keys if you want to change the displayed relative mV value.

• Press CFM to confirm the relative mV value. The relative mV offset is

displayed on the primary LCD and “OFF” message on the secondary LCD for

a few seconds. The instrument returns to measurement mode.

Note: The relative mV value can be changed only inside the relative mV offset

window (± 2000 mV).