Data transmission from the instrument to the PC can be done with the HI

92000 Windows

®

compatible software (optional). HI 92000 also offers

graphing and on-line help feature.

Data can be exported to the most popular spreadsheet programs for further

analysis.

To connect your instrument to a PC, use the optional Hanna HI 920010 cable

connector. Make sure that your instrument is switched off and plug one

connector to the instrument RS232C socket and the other into the serial port

of your PC.

Note: • Other cables than HI 920010 may use a different configuration.

In this case communication between instrument and PC may not

be possible.

• If you are not using Hanna Instruments HI 92000 software,

please see the following instructions.

SENDING COMMANDS FROM PC

It is also possible to remotely control the instrument with any terminal

program. Use HI 920010 cable to connect the instrument to a PC, start the

terminal program and set the communication options as follows: 8, N, 1, no

flow control.
COMMAND TYPES
To send a command to the instrument the scheme is:

<command prefix> <command> <CR>

where:

<command prefix> is a selectable ASCII character

between 0 and 47.

<command> is the command code (3 characters).

Note: Either small or capital letters can be used.
SIMPLE COMMANDS
PHR

Sets the range to pH 0.001 resolution

MVR

Sets the range to mV (HI 113 only)

MRS

Sets the range to relative mV (HI 113 only)

RNG

Is equivalent to pressing RANGE

CAL

Is equivalent to pressing CAL

CFM

Is equivalent to pressing CFM/GLP

UPC

Is equivalent to pressing the UP arrow key

DWC

Is equivalent to pressing the DOWN arrow key

MEM

Is equivalent to pressing MEM

MRR

Is equivalent to pressing MR

SET

Is equivalent to pressing SET

PC INTERFACE (HI 112 AND HI 113)