Safety, Areas of use, Storage, handling and transit – Heath Consultants Gasurveyor 500 Series User Manual

Page 4: Handling, Storage, Transit

Advertising
background image

GASURVEYOR 3-500 USER HANDBOOK

ii

SAFETY

The instrument must be regularly serviced and calibrated by fully trained
personnel in a safe area.

Batteries: Alkaline batteries or *Rechargeable battery pack must be
exchanged (*and recharged) in a safe area and fitted correctly before use.
Never use damaged batteries or expose to extreme heat.
See Section 4 : OPERATOR MAINTENANCE.

Only GMI replacement parts should be used.

If the instrument detects gas, follow your own organisation’s procedures and
operational guidelines.

The combustion chamber is a flameproof assembly and must not be opened
in the presence of a flammable atmosphere.

Gasurveyor 3-500 instruments are certified as EEx iad IIC T4

(-20oC< Tamb < 50oC). BAS01ATEX2292

II 2 G.

UL Class 1 Groups A, B, C and D.

This equipment is designed and manufactured to protect against other
hazards as defined in paragraph 1.2.7 of Annex II of the ATEX Directive
94/9/EC

Any right of claim relating to product liability or consequential damage to any third
party against GMI is removed if the warnings are not observed.

AREAS OF USE

Exposure to certain chemicals can result in a loss of sensitivity of the flammable
sensor. Where such environments are known or suspected it is recommended that
more frequent response checks are carried out. The chemical compounds that can
cause loss of sensitivity include Silicones, Lead, Halogens and Sulphur. Do not
use instrument in potentially hazardous atmospheres containing greater than 21%
Oxygen. The enclosure material is polypropylene and must not be exposed to
environments which are liable to result in mechanical or thermal degradation or to
damage caused by contact with aggressive substances. Additional protection may
be required in environments where the instrument enclosure is liable to damage.

STORAGE, HANDLING AND TRANSIT

The batteries in the rechargeable pack contain considerable energy and care should
be taken in their handling and disposal. Battery packs should be removed if the
instrument is stored for longer than 3 months. The instrument is designed to handle
harsh environments. The sensing elements are sealed to IP54 and the rest of the
instrument to IP64. If not subject to misuse or malicious damage, the instrument
will provide many years of reliable service. The instrument contains electrochemical
sensors with a life of 2 years. Under conditions of prolonged storage the sensors
should be removed. The sensor contains potentially corrosive liquid and care should
be taken when handling or disposing of the sensor, particularly when a leak is
suspected.