GASURVEYOR 3-500 USER HANDBOOK

B-4

In CAL mode, the instrument buttons have the functions
shown in Figure B-4.

Single Press

Double Press

Press and

Hold

Button 1

Button 2

Figure B-4 CAL Mode Button Functions

Zeroing the Instrument

1) Enter FCM. See the previous section ENTERING

FCM.

2) Double press Button Two

to zero current gas

range.

3) Single press Button Two

to select the next gas

range.

4) Repeat steps 2 and 3 until all gas ranges have been

zeroed.

5) Proceed to FIELD CALIBRATION

PROCEDURE to calibrate the instrument.