Selectable ranges in fcm, Entering fcm – Heath Consultants Gasurveyor 500 Series User Manual
Page 51
B-3
Selectable Ranges in FCM
When in FCM the following ranges are manually
selectable by pressing Button Two
: LEL, Volume
GAS.
Entering FCM
1) Switch the instrument on and allow it to complete
its warm-up checks.
2) Double press Button One
to initiate instrument
switch off. While OFF is displayed in the LCD
and before the instrument actually switches off,
enter the access code.
Note: Allow at least one second between button presses
when entering the button sequence. The default
(factory set) entry code is button sequence 1,2,1,2.
Alternative codes are user selectable.
When the instrument is in FCM, the “CAL” message
alternates on the display with the currently selected range.
An example of the display is shown in Figure B-3.
Figure B-3 Field Calibration Display
FIELD CALIBRATION