Selectable Ranges in FCM

When in FCM the following ranges are manually

selectable by pressing Button Two

: LEL, Volume

GAS.

Entering FCM

1) Switch the instrument on and allow it to complete

its warm-up checks.

2) Double press Button One

to initiate instrument

switch off. While OFF is displayed in the LCD
and before the instrument actually switches off,
enter the access code.

Note: Allow at least one second between button presses

when entering the button sequence. The default
(factory set) entry code is button sequence 1,2,1,2.
Alternative codes are user selectable.

When the instrument is in FCM, the “CAL” message
alternates on the display with the currently selected range.
An example of the display is shown in Figure B-3.

Figure B-3 Field Calibration Display

FIELD CALIBRATION