Heath Consultants Odorator User Manual

Page 12

Advertising
background image

O

peration

12. If the exhaust does not have a readily detectable odor repeat step 11. If

the exhaust does have a readily detectable odor, push and hold the read

switch down and record the percent gas in air by volume reading as

shown on the LCD. This reading is the readily detectable level which

should be noticed by the average person’s sense of smell.

13. Close the gas supply outlet valve, the flow metering valve and then

disconnect the sample hose from the gas supply outlet.

14. Leave the ODORATOR “ON” for approximately one minute after the

test to purge the mixing chamber.

15. Turn the ODORATOR “OFF”.

16. If chosen, correct the reading value using the correction chart. The

correction chart is included to linearize the instrument in its range.

To use the chart, take the ODORATOR readily detectable level

reading and look up and record the corrected actual % gas value.

For ODORATOR readings not specified within the chart, linearly

interpolate between the next highest and lowest points on the chart. Do

not extrapolate beyond the readings in the chart. The readily detectable

level reading and, if chosen, the corrected actual value must meet

the requirements of the user’s company operations procedures.

Use at Elevation. See Appendix C, page 11.

3

Advertising