Heath Consultants Odorator User Manual

Page 20

Advertising
background image

Appendix C

11

Appendix C

ODORATOR Use at Elevation

The ODORATOR works by allowing a metered amount of sample gas to

mix with a fixed amount of air and the mixture is then tested by a user for

odorant detection. The internal components of the ODORATOR are very

stable over changes in elevation. The air that is mixed, however, thins out

with increasing elevation and must be accounted for. An elevation change

of 3000 feet from the calibration can produce reading discrepancies of nearly

10 % which may be corrected. It is recommended that the ODORATOR

readings be corrected whenever the elevation change from calibration is

1000 feet or more. The calibration elevation may be found on the exhaust

port label.

A correction chart with accompanying usage directions has been developed.

The chart provides corrections for both increased and decreased elevations.

Most instruments will be used at elevations higher than their calibration

elevation. ODORATORS calibrated at the factory have a calibration el-

evation of about 50 feet. Many gas utilities send their ODORATORS to

the Heath factory for periodic calibration, hence most instruments will be

used at elevations higher than their calibration elevation. ODORATORS

calibrated at one elevation and then used at a higher elevation will make the

most use of this chart. An ODORATOR calibrated and used within 1000

feet of the same elevation will use the respective correction chart found on

the instrument.

The elevation correction chart is based on a normal atmosphere. Abnor-

mal atmospheres, as caused by extreme weather conditions, can affect

the measurements and safe use of the ODORATOR and operation is not

recommended. Usage instructions are included with the chart but use the

difference in elevation between the working or measurement elevation and

the calibration elevation to select the respective correction value versus the

ODORATOR’S reading.

Advertising