Heath Consultants Odorator User Manual

Page 5

Advertising
background image

i

Natural gas (methane) to be tested enters the ODORATOR through

the barbed fitting at the gas supply inlet port. Internally, the gas passes

through an internal five micron filter to the input of a low pressure

regulator. It then passes from the output of the low pressure regulator

to the flow metering valve. This valve, located on the top panel of

the ODORATOR, is operated by the user. After passing through the

valve (when opened), the gas enters the mass flow sensor. This is

the heart of the ODORATOR. The mass flow sensor responds to gas

flow within the range of the flow metering valve.

The signal from the mass flow sensor is then processed, converted to

a digital signal and fed to a liquid crystal display (LCD). The gas from

the mass flow sensor then passes to the mixing chamber where it is

combined with air drawn in by a blower wheel. The user then sniffs

the gas / air mixture at the exhaust port. The flow metering valve

is incrementally opened and the exhaust port sniffed until odorant

is readily detectable at which time the user presses and holds the

READ switch down to display percent gas in air by volume.

The electronics and the speed of the blower wheel are calibrated so

that the concentration of the methane gas / air sample at the exhaust

port agrees with the LCD at the 20% Lower Explosive Level (L.E.L.)

which is 1.00 % gas. All other readings may be corrected using

the correction chart. The correction chart is included to linearize

the instrument within its range. To use the chart, record both the

ODORATOR readily detectable level reading and the corrected actual

value. For ODORATOR readings not specified in the chart, linearly

interpolate between the next highest and lowest points on the chart.

Do not

extrapolate

beyond the readings in the chart. Depending on

how much the flow metering valve is opened, the concentration of gas

in air at the exhaust port will be anywhere from zero to approximately

2%. The LCD can only indicate up to 1.99%. Concentrations above

1.99% will be indicated by the LCD having its last two digits dashed

out as “1.--”.

INTRODUCTION