Heath Consultants Odorator User Manual

Page 22

Advertising
background image

Appendix C

13

Elevation Corrections for the Methane ODORATOR

The methane ODORATOR mixes air with natural gas to allow the user to

determine the percent gas level at which odorant is readily detectable. As

elevation is increased or decreased, the air density changes and must be

accounted for. Use the Elevation Change Correction Chart as follows:

1. Use the ODORATOR to determine the readily detectable level per

the instructions in this manual.

2. Write down the readily detectable reading from the LCD on the

ODORATOR.

3. Write down the elevation at which the ODORATOR was last

calibrated as found on the label on the inside of the exhaust port

cover.

4. Determine the elevation at which the reading was taken.

5. Subtract the calibration elevation from the reading elevation and

round off to the nearest 1000’.

6. Positive elevation differences correspond to the “UP” columns in

the chart while negative elevation differences correspond to the

“DOWN” colunms. Determine which elevation change correction

chart column to be used.

7. Use the ODORATOR reading and either look up the actual % gas

value from the chosen column or interpolate the actual % gas value

for the readings between the values on the chart and write the actual

% gas value down.

Interpolation Example 1: UP Column

An ODORATOR reading of .23 is taken at 3000’. The calibration elevation

is 50’ so the elevation difference is 3000 - 50 = 2950, which is rounded

off to 3000’. The difference is positive so the UP 3000 correction column

will be used.

A reading of .23 lies between chart readings of .20 and .30. This reading

also lies between actual values of .19 and .28 using the UP 3000 correction

column.

Advertising