Pressure display module (pdm), Tachometer display module (tdm), 4848b reactor controller – Parr Instrument 4848B User Manual

Page 16: Pressure

Advertising
background image

4848B Reactor Controller

P a r r I n s t r u m e n t C o m p a n y

14

Pressure Display Module (PDM)

PRESSURE

When this optional module is present, the left dis-
play in red shows the pressure, and the right hand
display in green shows the Pressure Alarm setpoint.
If the reactor pressure exceeds this setpoint, the
pressure module will trip the High Limit Alarm, cut-
ting power to the heater.

This module provides a digital read out for continu-
ously monitoring the pressure within the reactor,
plus a backup safety feature which will terminate
power to the heater if a preset maximum is reached.
Pressure is displayed with a 1 psi resolution and a
1% FS accuracy. The safety cut-out feature offers
excellent protection against accidental over-pressure
by allowing the user to set a maximum pressure
which, if reached, will activate the high limit relay
and turn off the heater immediately.

The signal for the pressure display is generated by
a transducer in a stainless steel housing mounted
on the reactor. The transducer must not be heated
above 120 °C; an elevated temperature can caused
the accuracy of the transducer to suffer. A water
cooling sleeve attached to the transducer mounting
body can be used to keep the transducer below this
limit when operating the reactor at higher tempera-
tures.

For applications involving reactants corrosive to
stainless steel, the transducer body can be replaced
with an isolator of special alloys which will keep the
reactants away from the transducer. Users should
consult with the Parr Technical Support before pur-
chasing one of these.

The Pressure Alarm setpoint may be adjusted by do-
ing the following:

1. Press the up or down arrow until the lower

display shows the desired setpoint. The lower
display will blink, showing the value has not
been set yet.

2. Press the set button. The lower display will stop

blinking, showing the new value has been set.

When ordering a pressure display module sepa-
rately or installed in any Parr controller, always
specify the reactor on which the transducer will be
installed so that the proper mounting hardware can
be furnished.

Note:

if the pressure transducer is not attached

to the 4848B Controller, the High Limit Reset

may trip or show a fi xed but incorrect value.

We recommend always operating with the

transducer attached.

Tachometer Display Module (TDM)

RPM

This optional feature can be included in any 4848B
Controller to provide a means for continuously
monitoring the stirring speed in a Parr reactor. It
consists of an optical sensor installed in the overarm
drive on the reactor which is then connected to the
4848B Controller.

The connecting cable has a plug which fi ts the
marked socket on the back panel of the controller.
The digital display will show the stirring speed over
a range of 0-2200 rpm with a 1 rpm resolution and
+/- 10 rpm accuracy.

Note:

The SV display serves no function on a

TDM

Users wishing to control the stirring speed remotely
with setpoint control ought to consider the Motor
Control Module (MCM) in lieu of the TDM.