General specifi cations, Environmental conditions, Provisions for lifting and carrying – Parr Instrument 4848B User Manual

Page 6: 4848b reactor controller specifi cations, 4848b reactor controller

Advertising
background image

4848B Reactor Controller

P a r r I n s t r u m e n t C o m p a n y

4

General Specifi cations

Electrical Ratings
115VAC, 15.0 Amps, 50/60 Hz

230VAC, 15.0 Amps, 50/60 Hz

Before connecting a controller to an electrical outlet,
the user must be certain that the electrical outlet has
an earth ground connection and that the line, load
and other characteristics of the installation do not
exceed the following limits:

Voltage: Fluctuations in the line voltage should not
exceed 10% of the rated nominal voltage shown on
the data plate.

Frequency: Controllers can be operated from either
a 50 or 60 Hertz power supply without affecting their
operation or calibration.

Current: The total current drawn should not exceed
the rating shown on the data plate on the controller
by more than 10 percent.

Thermocouple: Unless otherwise specifi ed, all
4848B Controllers operate with a Type-J (Iron-Con-
stantan) thermocouple. The total resistance of the
thermocouple and the lead wires should not exceed
100 ohms. If the resistance of the thermocouple
circuit is higher, it will not function properly.

Environmental Conditions

This instrument is intended to be used indoors.

Operating: 15 ºC to 35 ºC; maximum relative humid-
ity of 80% non-condensing. Installation Category II
(over voltage) in accordance with IEC 664.
Pollution degree 2 in accordance with IEC 664.

Altitude Limit: 2,000 meters.

Storage: -25 ºC and 65 ºC; 10% to 85% relative hu-
midity.

Provisions for Lifting and Carrying

Before moving the instrument, disconnect all con-
nections from the rear of the apparatus. Lift the
instrument by grabbing underneath each corner.

Caution!

Do not use in hazardous atmospheres.

4848B Reactor Controller Specifi cations

Input Thermocouple
type

Type-J

Operating Range

0-800 °C

Readout and Setpoint
Resolution

1 °C

System Accuracy

+/-2 °C

Control action

three-term PID plus limit
control

Control output 1

115VAC or 230VAC

Control output 2

.2 amp max
115VAC or 230VAC

Operating voltages

115VAC or 230 VAC

Heater Loop

Proportional Band:

12 C

Integral time:

375 sec

Derivative:

93 C/sec

Cycle time:

20 sec

Cooling Loop

Proportional Band:

5 C

Cycle time:

5 sec