4848b factory defaults, Navigating the menu system/factorydefault settings, Primary temperature module – Parr Instrument 4848B User Manual

Page 22: 4848b reactor controller

Advertising
background image

4848B Reactor Controller

P a r r I n s t r u m e n t C o m p a n y

20

4848B Factory Defaults

Navigating the Menu System/Factory

Default Settings

There are three broad categories of functions in the
4848B Controller modules. They are Regulation, Op-
eration, and Initial Settings. From the main screen,
Regulation mode is accessed by pressing the set
button. The Initial Setting mode is accessed by
holding the set button for three seconds. Operation
mode is accessed by pressing the circular arrow.

In each mode, you can cycle through the variables
by pressing the circular arrow. To change a value,
use the up or down arrow, which causes the value
to fl ash. Then press the set button to solidify the
choice. Return to the main screen by pressing set
again.

Keys command:
1. Press “SET” to select
2. Press return key move to next operation mode
3. Up/Down arrow keys to adjust value or select

type

Primary Temperature Module

Press return key and release

Operation

Mode

Select

type/value

Comment

r-S

Run

Run/Stop

SP

0**

Decimal point position

AL1H

375

Upper alarm setting

LoC

OFF

Lock mode (lock all keys or
only up/down arrow able to
use)

Out1

- (read-only) Heater output %

Out2

- (read-only) Cooling output %

** If using a type-J or type-K thermocouple, SP = 0

If using an RTD, SP = 1

Press and hold down “SET” for 5-sec

Operation

Mode

Select

type/value Comment

InPt

J**

Input type

tPUn

C

Temperature unit

tP-H

800**

Upper-limit range

tP-L

0

Lower-limit temperature range

CTRL

PID

Control mode (ON/OFF, MANU-
AL, PID and PID PROG)

S-HC

H1C2

Output 1 is Heating and output
2 is Cooling

ALA1

6

Absolute upper limit alarm op-
erates when PV value is higher
than AL1H setting

ALA2

0

Alarm 2 is disabled

SALA

OFF

System alarm

CoSH

ON

Communication write function
that able to use set point from
software

C-Sl

RTU

Format type

C-no

*

Communication address

bPS

9600

Communication baud rate

Len

8

Data length setting

PrtY

Even

Parity bit setting

StoP

1

Stop bit setting

* See note on address at end of Factory Defaults
** If using a type-J thermocouple, InPt = J, tP-H = 800

If using a type-K thermocouple, InPt = K, tP-H = 800

If using an RTD, InPt = Pt, tP-H = 600

Press “SET” and release

Operation

Mode

Select

type/value Comment

At

OFF

Auto Tuning ON/OFF

PID0

0

The 0th PID Parameter

SV

0

The 0th SV

P0

11.8

Proportional control

I0

375

Integral control

D0

93

Derivative control

IoF0

0

Integral value offset

HtPd

5

Heat cycle control

HcPd

5

Cool cycle control

CoEF

1

1 & 2 value output group dur-
ing dual loop output control

dEAd

0

Dead band

tPoF

*

PV Offset

* Set at factory