Tdm / mcm module, Pressure module, 4848b reactor controller – Parr Instrument 4848B User Manual

Page 23

Advertising
background image

4848B Reactor Controller

w w w . p a r r i n s t . c o m

21

TDM / MCM Module

Press return key and release

Operation

Mode

Select

type/value Comment

r-S

Run

Run/Stop

SP

0

Decimal point position

LoC

OFF

Lock mode (lock all keys or
only up/down arrow able to
use)

Out1

- (read-only)

Motor output %

Press and hold down “SET” for 5-sec

Operation

Mode

Select

type/value Comment

InPt

v5

Input type (v5 = 0-5V, v10 =
0-10V, nA0 = 0-20mA, nA4 =
4-20mA, nV =0-50mV)

tP-H

2206

Upper-limit range

tP-L

0

Lower-limit temperature range

CTRL

PID

Control mode (ON/OFF, MAN-
UAL, PID and PID PROG)

S-HC

Heat

Heat/Cool control

ALA1

0

Alarm 1 is disabled

SALA

OFF

System alarm

CoSH

ON

Communication write func-
tion that able to use set point
from software

C-Sl

RTU

Format type

C-no

*

Communication address

bPS

9600

Communication baud rate

Len

8

Data length setting

PrtY

Even

Parity bit setting

StoP

1

Stop bit setting

* See note on address at end of Factory Defaults

Press “SET” and release

Operation

Mode

Select

type/value Comment

At

OFF

Auto Tuning ON/OFF

PID0

0

The 0th PID Parameter

SV

0

The 0th SV

P0

170

Proportional control

C0

1

Integral control

D0

0

Derivative control

CoF0

0

Integral value

HtPd

.5

Heat/Cool cycle control

tPoF

*

PV Offset

CrHI

0

Analog output high

CrLo

0

Analog output low

* Set at factory

Pressure Module

Press return key and release

Operation

Mode

Select

type/value

Comment

r-S

Run

Run/Stop

SP

To be deter-
mined**

Decimal point position

AL1H

0

Upper-limit alarm 1

LoC

OFF

Lock mode (lock all keys or
only up/down arrow able to
use)

** If units are psi, SP = 0

If units are Bar, SP = 1

If units are MPa, SP = 2

Press and hold down “SET” for 5-sec

Operation

Mode

Select

type/value

Comment

InPt

V5

Input type (V5 = 0-5V, V10 =
0-10V, nA0 = 0-20mA, nA4 =
4-20mA, nV =0-50mV)

tP-H

Determined
by transduc-
er**

Upper-limit pressure range

tP-L

0
[-14.7 if com-
pound psi]

[-1.0 if com-
pound bar]

[-0.1 if com-
pound mPa]

Lower-limit pressure range

CTRL

ON/OFF

Control mode (ON/OFF,
MANUAL, PID and PID-
PROG)

S-HC

Heat

Heat/Cool control

ALA1

2

Alarm operates when PV
value is higher than SV
Value + ALH setting

SALA

OFF

System alarm

CoSH

ON

Communication write func-
tion that able to use set
point from software

C-Sl

RTU

Format type

C-no

*

Communication address

bPS

9600

Communication baud rate

Len

8

Data length setting

PrtY

Even

Parity bit setting

StoP

1

Stop bit setting

* See note on address at end of Factory Defaults