Lamp assembly installation – UVP 81-0333-01 BioLite Xe User Manual

Page 9

Advertising
background image

BioLite Xe MultiSpectral Light Source

9

7. Remove the rubber cap from the input side of the fiber optic trunk and plug the cable into the left

port on the BioLite unit.


Fiber Optic Light Table (Transilluminator) Installation

1. Remove one stopper at the left side of the darkroom.

2. If the imaging system already contains a transilluminator, unplug and remove the transilluminator.

3. Insert the fiber optic light table transilluminator in the imaging system.

4. Thread the knurled (non-smooth side) of the fiber optic cable from the inside of the darkroom

through the hole created by removing the stopper until the large bulb separating the smooth and
rough side of the fiber optic cable can no longer be pushed through the opening.

5. Remove the rubber cap from the input side of the fiber optic trunk and plug the cable into the left

port on the BioLite unit.

Lamp Assembly Installation

The BioLite Xe is shipped from the factory with the xenon lamp assembly uninstalled. Prior to initial use, the
lamp assembly must be installed in the BioLite Xe as described below.

The BioLite Xe xenon light bulb is a lamp assembly which includes a mounting plate and bracket. The
assembly is installed through the bottom of the unit as shown:

1. Unplug the power cable and USB cable from the back of the BioLIte Xe.

Plug the Fiber

Optic Cable into

the left port

Mounting Plate

Bracket

Light Bulb