5.

FAULT FINDING

Typical problems are shown in the following table.

Symptom

Possible Cause

Service Hint

Pump does not switch

ON or run (Green LED

not lit).

Battery fully discharged or

unserviceable.

Recharge for 14 hours.

Wiring loom disconnected

at some point / Fault on

joint.

Locate & resolder.

Circuit / Keypad failure.

Return for repair.

Pump does not run

(Green LED lit).

Pump or motor seizure.

Replace pump / motor.

Broken wire or worn

brushes.

Repair connection.

Pump runs but no air

drawn.

Diaphragm rubber split.

Replace.

Tubes between pump

and outlet split or

disconnected.

Replace.

Pump runs but cannot

achieve desired flow

rate.

Leak on pump assembly. Locate & Correct.
Dirty or damaged valves.

Clean / Replace.

Inlet filter clogged.

Remove & Clean.

Excessive pressure drop. Check sample filter.

Red error LED lit or

flashing.

Battery voltage too low.

Recharge battery.

Excessive pressure drop

or high pump current.

Remove pressure load

/ Repair pump.

Flow rate not main-

tained through full

working shift.

Incomplete charge.

Re-charge.

Battery below capacity.

Fully charge & check

voltage which should

be 13 - 15 V. If below

this range, check the

output from the

charger. It is typically

14 V at full charge. If

this is correct, replace

the battery.

High current consumption

due to mechanical defect

in pump.

Locate & correct.

Filter disc clogged.

Replace.

Calibration adjustment

required.

Adjust the “CAL”

control

Fault Finding

Page 10 of 12

VORTEX Ultra Flow & Ultra Flow lite

SAMPLING PUMPS Handbook