Casella CEL Vortex Ultraflow area sampling pump User Manual

Page 4

Advertising
background image

2.

OPERATION

Both designs utilise a twin diaphragm pump to reduce pulsations in airflow

and maintain high efficiency. Precise control of the sample flow rate is

achieved using a current proportional feed back technique.

Both pumps have similar front panels as shown below.

2.1

Battery

The battery is a sealed lead acid type, giving 7 Ah at 12 volts.

The run time is dependent upon flow and pressure conditions, for example:

Asbestos clearance

8 litres/min using a 0.8 µ, 25 mm filter the run time is typically

5 hours.

PM10/2.5

16.7 litres/min using a 47 mm GFA filter the run time is typically in

excess of 4 hours.

The charging socket is located at the back of the pump. Using the charger

provided, the battery can be fully charged in 12-14 hours. The three-stage

charger provided prevents overcharging.

To ensure maximum life expectancy of the battery pack, it is recommended

that the battery is not left in a fully discharged condition for extended periods

of time.

2.2

Battery Low Shutdown

A battery low cut-off facility will turn the pump off once the internal battery

pack becomes fully discharged. This will prevent “deep discharge” damage

Operation

Page 4 of 12

VORTEX Ultra Flow & Ultra Flow lite

SAMPLING PUMPS Handbook

Advertising