Casella CEL Vortex Ultraflow area sampling pump User Manual

Page 7

Advertising
background image

3.

Connect a clean filter to the pump and adjust to the required flow

rate using the front panel keys.

4.

With the pump running and the filter connected, adjust the

potentiometer anti-clockwise until the error LED flashes, then back

the adjustment off half a turn.

5.

Replace the cover.

2.8

Timer

The LCD timer display situated on the front panel indicates the sample run

elapsed time in whole minutes. The timer can be reset to zero by holding

down the “RESET” key for 3 seconds. This can be carried out while the

pump is on or off.

2.9

Table of Functionality

Function

Required Key

Press

LED

Buzzer

Switch ON

Press both keys

together.

Green “PUMP

ON” LED lit.

Double bleep.

Switch OFF

Press both keys

together.

No LEDs lit.

Flow adjustment

mode

Press “FLOW”

key for 3 seconds.

Green “PUMP

ON” LED flashes.

Exit flow adjust-

ment mode

No key press for 4

seconds.

Green “PUMP

ON” LED stops

flashing.

Reset timer

Press “RESET”

key for 3 seconds.

Operation

Pot VR2

The location of the potentiometer VR2

Page 7 of 12

VORTEX Ultra Flow & Ultra Flow lite
SAMPLING PUMPS Handbook