Prog. set flo oc l/min reset cal i.r – Casella CEL Sampling heads for personal dust sampling User Manual

Page 16

Advertising
background image

6.1.2 Calibration at High Flow Using a BIOS Dry Cal

Flowmeter

Attach the sampling head and Dry Cal Flowmeter to the inlet nozzle of an

Apex personal sampling pump to measure the actual flow as shown in Figure

13.

1.

Start the Apex unit and wait for it to enter Hold Mode

2.

Press

or

several times until FLOW SEt blinks on the

display.

3.

Press

to select it.

3.

Use

and

to set the flow required for calibration.

As a general rule, the pump should be calibrated to the flow required

for the proposed task.

4.

Press

to accept the setting. The pump will start.

5.

As soon as CAL SEt is displayed, use

and

to adjust the

flow being measured by the Dry Cal Flowmeter to match the

required calibration point.

6.

When the flow rate is acceptable, press

to accept the value and

return to Hold Mode.
This calibration will be saved and used by the Apex unit in all future

samples made at this flow.

Figure 13: Arrangement for calibration using a Dry Cal Flowmeter and Apex

Unit

PROG.

SET FLO

oC

L/min

RESET

CAL

I.R.

TEMP 19.48

2.5

02040

Clamp

Clamping

studs

Stand

Sampling

head

Connecting tube

(5 mm nominal internal dia.)

Flow

Flow

Outlet

Dry Cal Flowmeter

Apex Pump

Dry Cal

®

Primary Flow Meter

Calibration

Page 16 of 24

SAMPLING HEADS AND ACCESSORIES

Users Guide

Advertising