3.

TOTAL INHALABLE SAMPLING HEADS

3.1

I.O.M. Head (Part No. P109009A)

I.O.M. Cassette, (Part No. P109010)

The I.O.M. (Institute of Occupational Medicine)

inhalable dust sampler is now firmly

established both in Europe and America. It

employs a cassette system that eliminates

direct handling of the filter paper and closely

simulates the particle collection behaviour of

the nose and mouth. Passing the flow through

a cassette prevents particulate from settling on

the inside of the head. Because the whole

cassette is weighed, this offers the additional

benefit of minimising under- sampling.

The I.O.M. Head is a true inhalable

sampler as specified in MDHS 14/3 and the

ISO/CEN criteria. It allows the collection of

many different dust fractions, either

individually, or at the same time when using

PUF foams.

When polyurethane foam (PUF) plugs are used in the inlet, it is

possible to collect two different fractions of dust at the same time, thereby

saving valuable time and money. By using the correct combination of foams it

is possible to collect combinations of Inhalable, Thoracic or Respirable dust.

02045

Sampler body

Fat "O" Ring

Casette lower half

Filter

Air Flow

Air
Flow

Casette upper half

Thin "O" Ring

Front cover

Dust cap

Figure 1: The I.O.M. sampling head assembly

Dust & Aerosol Heads

Page 7 of 24

SAMPLING HEADS AND ACCESSORIES
Users Guide